Z-2168-2013Class II
OPERON D850 Surgical Table
Berchtold Corp. / initiated 2012-04-16 / Terminated / root cause: manufacturing process
Berchtold Corp. began a recall of OPERON D850 Surgical Table Product Usage: Surgical Table on . FDA classified it as Class II and gives the reason as "Table was not lowering". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2168-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Berchtold Corp.
- Product
- OPERON D850 Surgical Table Product Usage: Surgical Table
- Product code
- GDC Table, Operating-Room, Electrical
- Reason for recall
- Table was not lowering.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Berchtold sent a Customer Notification letter dated April 16, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers that BERCHTOLD will upgrade the OPERON Surgical tables with a new firmware version that will eliminate the potential for experiencing problems. For questions call 843-569-6100 ext. 322.
- Quantity in commerce
- 263
- Distribution
- USA Nationwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, HI.IL, IA, KY, LA, MD, MA, MI, MN, MS, NH, NJ, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WI and including Canada.
- Product codes and lots
- 3501110-S10023, 3501110-S10024, 3501110-S10025, 3501110-S10205, 3501110-S10208, 3501110-S10064, 3501110-S10036, 3501110-S10043, 3501110-S10045, 3501110-S10050, 3501110-S10061, 3501110-S10062, 3511110-S10139, 3511110-S10143, 3511110-S10144, 3511110-S10162, 3501110-S10192, 3501110-S10018, 3511110-S10038, 3511110-S10041, 3511110-S10047, 3511110-S10072, 3511110-S10137, 3511110-S10138, 3501110-S10153, 3501110-S10166, 3501110-S10093, 3501110-S10104, 3501110-S10090, 3501110-S10147, 3501110-S10262, 3501110-S10281, 3501110-S10105, 3501110-S10172, 3501110-S10195, 3501110-S10109, 3501110-S10134, 3501110-S10099, 3501110-S10101, 3501110-S10151, 3511110-S10086, 3511110-S10092, 3511110-S10085, 3511110-S10091, 3501110-S10065, 3501110-S10037, 3501110-S10056, 3501110-S10259, 3501110-S10083, 3501110-S10170, 3501110-S10212, 3501110-S10067, 3501110-S10073, 3501110-S10116, 3501110-S10117, 3501110-S10119, 3501110-S10157, 3501110-S10167, 3501110-S10204, 3501110-S10207, 3501110S-10110, 3511110-S10234, 3501110-S10078, 3501110-S10034, 3501110-S10035, 3501110-S10097, 3501110-S10112, 3501110-S10088, 3501110-S10096, 3501110-S10098, 3501110-S10102, 3501110-S10111, 3501110-S10019, 3501110-S10021, 3501110-S10022, 3501110-S10223, 3501110-S10267, 3501110-S10049, 3501110-S10044, 3501110-S10052, 3501110-S10149, 3501110-S10121, 3501110-S10129, 3501110-S10254, 3501110-S10257, 3501110-S10263, 3501110-S10100, 3511110-S10176, 3511110-S10221, 3501110-S10026, 3501110-S10051, 3501110-S10211, 3501110-S10251, 3501110-S10046, 3501110-S10053, 3501110-S10054, 3501110-S10055, 3501110-S10063, 3501110-S10080, 3501110-S10028, 3501110-S10027, 3501110-S10027, 3501110-S10152, 3511110-S10161, 3501110-S10079, 3501110-S10095, 3511110-S10039, 3501110-S10103, 3501110-S10209, 3501110-S10213, 3501110-S10076, 3501110-S10114, 350110-S10118, 3501110-S10126, 3501110-S10128, 3501110-S10136, 3501110-S10154, 3501110-S10156, 3501110-S10158, 3501110-S10258, 3501110-S10249, 3501110-S10256, 3501110-S10260, 3501110-S10261, 3501110-S10031, 3501110-S10210, 3501110-S10277, 3501110-S10070, 3501110-10082, 35011100-S10084, 3501110-S10087, 3501110-S10089, 3501110-S10248, 3501110-S10250, 3501110-S10131, 3501110-S10094, 3501110-S10016, 3501110-S10017, 3501110-S10020, 3501110-S10030, 3501110-S10206, 3501110-S10252, 3501110-S10148, 3501110-S10155, 3501110-S10171, 3501110-S10173, 3511110-S10107, 3501110-S10193, 35011110-S10233, 3511110-S10159, 3511110-S10160, 3511110-S10168, 3511110-S10169, 3511110-S10181, 3511110-S10196, 3511110-S10197, 3511110-S10198, 3511110-S10199, 3511110-S10200, 3511110-S10201, 3511110-S10202, 3511110-S10214, 3511110-S10215, 3511110-S10218, 3511110-S10219, 3511110-S10228, 3511110-S10235, 3501110-S10253, 3501110-S10255, 3511110-S10033, 3501110-S10074, 3501110-S10075, 3501110-S10077, 3511110-S10040, 3511110-S10145, 3511110-S10032, 3501110-S10270, 3501110-S10135, 3501110-S10150, 3501110-S10164, 3501110-S10165, 3501110-S10265, 3501110-S10266, 3501110-S10280, 3501110-S10060, 3501110-S10133, 3501110-S10106, 3511110-S10108, 3511110-S10217, 3511110-S10220, 3501110-S10113, 3501110-S10115, 3501110-S10120, 3501110-S10122, 3501110-S10123, 3501110-S10125, 3501110-S10130, 3501110-S10132, 3511110-S10140, 3511110-S10141, 3511110-S10142, 3511110-S10146, 3501110-S10066, 3501110-S10068, 3501110-S10069, 3501110-S10081, 3501110-S10222, 3501110-S10224, 3501110-S10225, 3501110-S10226, 3501110-S10227, 3501110-S10230, 3501110-S10231, 3501110-S10236, 3501110-S10237, 3501110-S10238, 3501110-S10239, 3501110-S10071, 3501110-S10124, 3501110-S10127, 3511110-S10177, 3511110-S10178, 3511110-S10179, 3511110-S10180, 3511110-S10203
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2168-2013 | Class II | Terminated | Voluntary: Firm initiated |