"***Flo Trac Sensor Model MHD85***" Sterile, single use kit that monitors pressures when attached to pressure...
Edwards Lifesciences, LLC began a recall of "***Flo Trac Sensor Model MHD85***" Sterile, single use kit that monitors pressures when attached to pressure monitoring catheters on . FDA classified it as Class II and gives the reason as "The Edwards Lifesciences FloTrac Sensor is being recalled due to a manufacturing non-conformance related to incorrect packaging configuration where the IFU was placed on top of the Tyvek instead of being placed at the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2189-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- "***Flo Trac Sensor Model MHD85***" Sterile, single use kit that monitors pressures when attached to pressure monitoring catheters.
- Product code
- DRS Transducer, Blood-Pressure, Extravascular
- Reason for recall
- The Edwards Lifesciences FloTrac Sensor is being recalled due to a manufacturing non-conformance related to incorrect packaging configuration where the IFU was placed on top of the Tyvek instead of being placed at the bottom of the box. There is no evidence available at this time that could validate the sterility of this incorrect packaging configuration.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Edwards Lifesciences initiated a US recall via an Urgent Product Recall letter dated July 31, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised that an Edweards Lifesciences representative would come to their facility to retrieve the affected device. A device replacement would be provided to the customer. Customers were asked to complete the enclosed Confirmation Form and give it to the Edwards Lifesciences representative that will come to their facility. For customers with questions were instruct4ed to call Edwards Customers Service at 1-800-424-3278. For questions regarding this recall call 972-410-7100.
- Quantity in commerce
- 35 units
- Distribution
- Nationwide Distribution including MO and MI.
- Product codes and lots
- 595531157
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2189-2013 | Class II | Terminated | Voluntary: Firm initiated |