Z-2166-2013Class II

Outlook Safety Infusion System Burette Set. For intravenous fluid administration

B Braun Medical Inc / initiated 2013-08-05 / Terminated / root cause: manufacturing process

B. Braun Medical, Inc. began a recall of Outlook Safety Infusion System Burette Set. For intravenous fluid administration on . FDA classified it as Class II and gives the reason as "Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2166-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
Outlook Safety Infusion System Burette Set. For intravenous fluid administration.
Product code
FRN Pump, Infusion
Reason for recall
Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On August 4, 2013, a letter was sent to direct accounts asking them to check their inventory for the affected product. These customers were also instructed to notify their customers if the product was further distributed and to return all affected product and/or the product removal acknowledgement form.
Quantity in commerce
900
Distribution
Distributed in the states of FL, NY, NJ, and TX.
Product codes and lots
Catelog Number 375038, Lot Number 0061317318, Expiration Date: 04/30/2016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2166-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2166-2013Class IITerminatedVoluntary: Firm initiated