Z-2276-2013Class II

Monaco RTP System Radiation treatment planning

Elekta Limited / initiated 2013-07-02 / Terminated / root cause: design

Elekta, Inc. began a recall of Monaco RTP System Radiation treatment planning on . FDA classified it as Class II and gives the reason as "Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2276-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco RTP System Radiation treatment planning
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.
Root cause tag
design
FDA root cause
Device Design
Action
Monaco 3.3 upgrades have been temporarily halted and Important Safety Notice (LINMON0007) has been issued to effected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that couch files are checked for releases 3.20.00 and 3.30.00 to verify that the correct relative electron density values are being used in the dose calculation. If customers have edited the sample couch files, and previously upgraded to Monaco 3.20.00 or 3.30.00, their couch files have already been overwritten. Open the couch files, make the necessary updates to the couch parameters, and use the "Save As Treatment Couch" option to save the modified sample couch files to a new, unique name. Please note that the couch files can also be edited in Focal and then the patients opened in Monaco. Comparable Focal releases would be Focal4.64.00(Monaco 3.10.00), Focal4.70.00(Monaco 3.20.00) and Focal4.80.00 (Monaco 3.30.00). For questions regarding this recall call 404-993-5886.
Quantity in commerce
45
Distribution
Nationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
Product codes and lots
Versions 3.10.00 and later

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2276-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2276-2013Class IITerminatedVoluntary: Firm initiated