Z-2276-2013Class II
Monaco RTP System Radiation treatment planning
Elekta Limited / initiated 2013-07-02 / Terminated / root cause: design
Elekta, Inc. began a recall of Monaco RTP System Radiation treatment planning on . FDA classified it as Class II and gives the reason as "Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2276-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco RTP System Radiation treatment planning
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Monaco 3.3 upgrades have been temporarily halted and Important Safety Notice (LINMON0007) has been issued to effected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that couch files are checked for releases 3.20.00 and 3.30.00 to verify that the correct relative electron density values are being used in the dose calculation. If customers have edited the sample couch files, and previously upgraded to Monaco 3.20.00 or 3.30.00, their couch files have already been overwritten. Open the couch files, make the necessary updates to the couch parameters, and use the "Save As Treatment Couch" option to save the modified sample couch files to a new, unique name. Please note that the couch files can also be edited in Focal and then the patients opened in Monaco. Comparable Focal releases would be Focal4.64.00(Monaco 3.10.00), Focal4.70.00(Monaco 3.20.00) and Focal4.80.00 (Monaco 3.30.00). For questions regarding this recall call 404-993-5886.
- Quantity in commerce
- 45
- Distribution
- Nationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
- Product codes and lots
- Versions 3.10.00 and later
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2276-2013 | Class II | Terminated | Voluntary: Firm initiated |