Z-2200-2013Class II

T8 Removal Head, Ref 8910641, Qty: 1 Ea

Medtronic Sofamor Danek USA Inc / initiated 2013-08-12 / Terminated / root cause: manufacturing process

Medtronic Sofamor Danek USA Inc began a recall of T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with... on . FDA classified it as Class II and gives the reason as "One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2200-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.
Product code
HXX Screwdriver
Reason for recall
One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medtronic sent an Urgent Voluntary Market Removal Recall letter dated August 9, 2013 with Recall Questionnaires and Confirmation forms included to Risk Managers and Sales Representatives. The letter identified the affected product, problem and action to be taken. Customers were instructed to remove the affected product, discontinue use and contact your Medtronic Sales Representative to coordinate the return and replenishment of affected product. For questions call your Medtronic Sales Representative or Global Quality Department at 800-876-3133.
Quantity in commerce
9 units
Distribution
USA Nationwide Distribution in the state of GA, AK, WA, and CA.
Product codes and lots
RS12K008

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2200-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2200-2013Class IITerminatedVoluntary: Firm initiated