510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2232-2013Class II

TWINFIX Ultra HA 6.5 mm with 2 ULTRABRAID Suture, Blue / Black; Part Number: 72202631 Biodegradable suture...

Smith & Nephew, Inc. / initiated 2013-08-06 / Terminated / root cause: packaging

Smith & Nephew, Inc. Endoscopy Division began a recall of TWINFIX Ultra HA 6.5 mm with 2 ULTRABRAID Suture, Blue / Black; Part Number: 72202631 Biodegradable suture anchor on . FDA classified it as Class II and gives the reason as "Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2232-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
TWINFIX Ultra HA 6.5 mm with 2 ULTRABRAID Suture, Blue / Black; Part Number: 72202631 Biodegradable suture anchor
Product code
MAI Fastener, Fixation, Biodegradable, Soft Tissue
Reason for recall
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Root cause tag
packaging
FDA root cause
Packaging
Action
Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Quantity in commerce
708 US
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
Product codes and lots
50325165 50326838 50336722 50339646 50339844 50343443 50350417 50352981 50379206 50397701 50429644

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2232-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2215-2013Class IITerminatedVoluntary: Firm initiated
Z-2216-2013Class IITerminatedVoluntary: Firm initiated
Z-2217-2013Class IITerminatedVoluntary: Firm initiated
Z-2218-2013Class IITerminatedVoluntary: Firm initiated
Z-2219-2013Class IITerminatedVoluntary: Firm initiated
Z-2220-2013Class IITerminatedVoluntary: Firm initiated
Z-2221-2013Class IITerminatedVoluntary: Firm initiated
Z-2222-2013Class IITerminatedVoluntary: Firm initiated
Z-2223-2013Class IITerminatedVoluntary: Firm initiated
Z-2224-2013Class IITerminatedVoluntary: Firm initiated
Z-2225-2013Class IITerminatedVoluntary: Firm initiated
Z-2226-2013Class IITerminatedVoluntary: Firm initiated
Z-2227-2013Class IITerminatedVoluntary: Firm initiated
Z-2228-2013Class IITerminatedVoluntary: Firm initiated
Z-2229-2013Class IITerminatedVoluntary: Firm initiated
Z-2230-2013Class IITerminatedVoluntary: Firm initiated
Z-2231-2013Class IITerminatedVoluntary: Firm initiated
Z-2232-2013Class IITerminatedVoluntary: Firm initiated
Z-2233-2013Class IITerminatedVoluntary: Firm initiated
Z-2234-2013Class IITerminatedVoluntary: Firm initiated
Z-2235-2013Class IITerminatedVoluntary: Firm initiated
Z-2236-2013Class IITerminatedVoluntary: Firm initiated
Z-2237-2013Class IITerminatedVoluntary: Firm initiated
Z-2238-2013Class IITerminatedVoluntary: Firm initiated
Z-2239-2013Class IITerminatedVoluntary: Firm initiated
Z-2240-2013Class IITerminatedVoluntary: Firm initiated