TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with...
Smith & Nephew, Inc. Endoscopy Division began a recall of TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with Needles; Part Number: 72202612 Biodegradable suture anchor on . FDA classified it as Class II and gives the reason as "Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2238-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TWINFIX" Ultra 4.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID") with Needles; Part Number: 72202612 Biodegradable suture anchor
- Product code
- MAI Fastener, Fixation, Biodegradable, Soft Tissue
- Reason for recall
- Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
- Quantity in commerce
- 1571 US
- Distribution
- Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
- Product codes and lots
- 50373723 50389654 50389655 50397577 50397596 50397599 50397605 50417226 50431744 50431853 50433695 50437528 50440608 50441690 50447116 50447118 50447119 50448227 50454009 50454867 50461412 50467503 50469159 50470653
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2215-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2216-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2217-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2218-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2219-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2220-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2221-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2222-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2223-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2224-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2225-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2226-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2227-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2228-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2229-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2230-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2231-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2232-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2233-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2234-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2235-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2236-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2237-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2238-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2239-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2240-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.