TWINFIX Ultra 6.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID) with Needles...
Smith & Nephew, Inc. Endoscopy Division began a recall of TWINFIX Ultra 6.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID) with Needles; Part Number: 72202620 Biodegradable suture anchor on . FDA classified it as Class II and gives the reason as "Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2240-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TWINFIX Ultra 6.5 mm PLLA-HA Suture Anchor with two #2 ULTRABRAID Sutures (blue, blue-COBRAID) with Needles; Part Number: 72202620 Biodegradable suture anchor
- Product code
- MAI Fastener, Fixation, Biodegradable, Soft Tissue
- Reason for recall
- Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
- Quantity in commerce
- 314 US
- Distribution
- Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
- Product codes and lots
- 50373721 50375642 50397595 50397598 50397604 50397607 50449548
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2215-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2216-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2217-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2218-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2219-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2220-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2221-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2222-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2223-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2224-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2225-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2226-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2227-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2228-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2229-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2230-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2231-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2232-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2233-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2234-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2235-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2236-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2237-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2238-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2239-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2240-2013 | Class II | Terminated | Voluntary: Firm initiated |