Z-2122-2013Class I

Microscan Synergies Plus Negative Urine Combo 2

Siemens Healthcare Diagnostics Inc. / initiated 2013-08-21 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Microscan Synergies Plus Negative Urine Combo 2 Product Usage: MicroScan Synergies plus Panels & MicroScan Rapids plus (Negative) - For use in determining... on . FDA classified it as Class I and gives the reason as "Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2122-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Microscan Synergies Plus Negative Urine Combo 2 Product Usage: MicroScan Synergies plus Panels & MicroScan Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens sent an Urgent Field Safety Notice letters dated August 2013 to all US customers via over night delivery, receipt requested. Letters for foreign customers are being translated and will be sent as soon as possible. The letter identified the affected product, reason for field correction, risk to health problem and actions to be taken by customer. Customers were instructed to suppress all reporting of sensitive and intermediate results for imipenem and meropenem. Siemens also recommends that the customer consider the need to review previous test results, conduct patient follow-up, and/or repeat testing. Customer were requested to complete the attached Field Correction Effectiveness Check form and fax it to 302-631-8467. For questions contact your Siemens Technical Support Provider or Distributor.
Quantity in commerce
1,324 boxes, 26,480 panels
Distribution
Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland.
Product codes and lots
Internal Number - B1025-112  Catalog Number - 10444749 Lot code/expiration date: 2012-07-12, 2012-07-12; 2012-10-03, 2012-10-03; 2013-06-08, 2013-06-08; 2013-12-17, 2013-12-17; 2014-04-22, 2014-04-22; 2014-06-11, 2014-06-11;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2122-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2119-2013Class ITerminatedVoluntary: Firm initiated
Z-2120-2013Class ITerminatedVoluntary: Firm initiated
Z-2121-2013Class ITerminatedVoluntary: Firm initiated
Z-2122-2013Class ITerminatedVoluntary: Firm initiated
Z-2123-2013Class ITerminatedVoluntary: Firm initiated
Z-2124-2013Class ITerminatedVoluntary: Firm initiated
Z-2125-2013Class ITerminatedVoluntary: Firm initiated
Z-2126-2013Class ITerminatedVoluntary: Firm initiated
Z-2127-2013Class ITerminatedVoluntary: Firm initiated
Z-2128-2013Class ITerminatedVoluntary: Firm initiated
Z-2129-2013Class ITerminatedVoluntary: Firm initiated