DRX- Revolution, Catalog number: 101937, 1023415 & 1023423. Made in U.S.A. by Carestream Health, Inc. 150...
Carestream Health Inc. began a recall of DRX- Revolution, Catalog number: 101937, 1023415 & 1023423. Made in U.S.A. by Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 to perform... on . FDA classified it as Class II and gives the reason as "Carestream Health Inc. has conducted a voluntary field action which includes the installation of a software patch on the following: DRX-Revolution Mobile X-Ray, DRX-Mobile and DRX Transportable Systems with DirectView...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0179-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- DRX- Revolution, Catalog number: 101937, 1023415 & 1023423. Made in U.S.A. by Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 to perform radiographic x-ray examinations on all pediatric and adult patients
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Carestream Health Inc. has conducted a voluntary field action which includes the installation of a software patch on the following: DRX-Revolution Mobile X-Ray, DRX-Mobile and DRX Transportable Systems with DirectView Version 5.5 and 5.6 console software when used with DRX-1 Detector version 69 firmware. Carestream recently received reports of a non-diagnostic image on exams when a Normal Exposure
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, Carestream Health Inc., sent an "Urgent: Medical Device Recall" letter dated August 29, 2013 to its customers. The letter described the product, problem and actions taken. Carestream has developed a software patch that will allow Normal Exposure Mode and Extended Exposure Mode images to be taken in any order and will install the patch in all DRX-Revolution, DRC-Mobile and DRX-Transportable Systems with software version 5.5 or higher and Detector firmware version 69 or higher. No DRX Transportables were affected. Your Carestream service representative will install the patch at your next service call. If you have questions or concerns, or if you experience an issue prior to our visit, please contact the Carestream Customer Care Center at 1-800-328-2910, 8am through 8pm Eastern Standard Time.
- Quantity in commerce
- 190 units
- Distribution
- Nationwide distribution:AL, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MI, MO, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT, VA, WA, WI, and WV.
- Product codes and lots
- Model: DRXR-1 Serial numbers: all serial numbers manufactured with version 5.5 software and version 69 firmware
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0179-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0180-2014 | Class II | Terminated | Voluntary: Firm initiated |