Z-0370-2014Class II
Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed...
Handicare USA, Inc. / initiated 2013-10-30 / Terminated / root cause: design
Ergosafe Products LLC DBA Prism Medical Services USA began a recall of Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed in the Operations Manual is Medcare Products... on . FDA classified it as Class II and gives the reason as "The lift might fail with the labeled 500-lb. load rating". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0370-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Handicare USA, Inc.
- Product
- Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed in the Operations Manual is Medcare Products, Burnsville, MN. Used for assisting patients/residents to a standing position.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- The lift might fail with the labeled 500-lb. load rating.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Medcare Products, sent an "URGENT FIELD CORRECTIVE ACTIONS" letter dated October 21, 2013 via certified mail on October 30, 2013 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to peel off the boom stickers and serial number label and replace with the new ones supplied; review the revised Manual and dispose of the original; and complete the action report form once the products have been relabeled. Please refer to your Operations Manual or visit our website at www.medcarelifts.com to view the manual online. If we can answer any questions that you may have regarding the testing standards or the relabeling of the products, please do not hesitate to contact us at 1-800-695-4479.
- Quantity in commerce
- 97 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- Serial numbers 0109STH0042, 0112STH0085, 0112STH0086, 0112STH0087, 0112STH0089, 0112STH0090, 0112STH0091, 0112STH0092, 1211STH0071, 0207STH0003, 0207STH0004, 0209STH0043, 0211STH0072, 0212STH0093, 0212STH0094, 0212STH0095, 0212STH0096, 0212STH0097, 0212STH0098, 0212STH0099, 0212STH0100, 0308STH0027, 0308STH0028, 0308STH0029, 0308STH0030, 0309STH0044, 0309STH0045, 0309STH0046, 0407STH0005, 0409STH0047, 0507STH0006, 0508STH0031, 0508STH0032, 0510STH0058, 0510STH0059, 0510STH0060, 0511STH0073, 0511STH0074, 0607STH0007, 0608STH0033, 0608STH0034, 0609STH0049, 0610STH0061, 0610STH0062, 0610STH0063, 0610STH0064, 0611STH0075, 0611STH0076, 0611STH0077, 0613STH0113, 0613STH0114, 0613STH0115, 0613STH0116, 0613STH0117, 0613STH0118, 0707STH0008, 0708STH0035, 0708STH0036, 0709STH0050, 0709STH0051, 0710STH0065, 0711STH0078, 0711STH0079, 0711STH0080, 0712STH0101, 0712STH0102, 0712STH0103, 0712STH0104, 0809STH0052, 0809STH0053, 0810STH0066, 0810STH0067, 0810STH0068, 0905STH0001, 0907STH0009, 0908STH0037, 0908STH0038, 0909STH0054, 0909STH0055, 0911STH0081, 0911STH0082, 1006STH0002, 1008STH0039, 1008STH0040, 1008STH0041, 1009STH0056, 1009STH0057, 1012STH0105, 1012STH0106, 1012STH0107, 1012STH0108, 1110STH0069, 1112STH0109, 1112STH0110, 1210STH0070, 1211STH0083, and 1211STH0084.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0370-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0371-2014 | Class II | Terminated | Voluntary: Firm initiated |