Z-0197-2014Class II

SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration...

Baxter Healthcare Corporation / initiated 2013-06-10 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0197-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.
Product code
FRN Pump, Infusion
Reason for recall
Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On June 10, 2013 Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Library. The notices were dated June 5, 2013 and were distributed via first class mail. Until further information is provided, users should continue to follow the SIGMA SPECTRUM Infusion Pump Operator's Manual, which identifies a number of environmental and clinical usage factors that contribute to upstream occlusion and air-in-line alarms. For high-risk patients, consider having additional devices as backup. If the biomedical engineer finds that the sensor is not functioning properly, contact Baxter Healthcare at 800-356-3454 (choose option 1). For clinical questions, please contact Medical Information Services at Baxter at 800-933-0303, Monday through Friday 8:00am to 5:00 pm Central Time.
Quantity in commerce
273,835 units
Distribution
Worldwide Distribution-USA including Puerto Rico and Canada.
Product codes and lots
Product Code: 35700BAX, 35700ABB Serial Numbers: All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0197-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0197-2014Class IITerminatedVoluntary: Firm initiated