SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration...
Baxter Healthcare Corp. began a recall of SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0197-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On June 10, 2013 Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Library. The notices were dated June 5, 2013 and were distributed via first class mail. Until further information is provided, users should continue to follow the SIGMA SPECTRUM Infusion Pump Operator's Manual, which identifies a number of environmental and clinical usage factors that contribute to upstream occlusion and air-in-line alarms. For high-risk patients, consider having additional devices as backup. If the biomedical engineer finds that the sensor is not functioning properly, contact Baxter Healthcare at 800-356-3454 (choose option 1). For clinical questions, please contact Medical Information Services at Baxter at 800-933-0303, Monday through Friday 8:00am to 5:00 pm Central Time.
- Quantity in commerce
- 273,835 units
- Distribution
- Worldwide Distribution-USA including Puerto Rico and Canada.
- Product codes and lots
- Product Code: 35700BAX, 35700ABB Serial Numbers: All
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0197-2014 | Class II | Terminated | Voluntary: Firm initiated |