DePuy Mitek FMS Outflow Tubing with One-Way valve
DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the... on . FDA classified it as Class II and gives the reason as "Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0192-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Mitek, Inc.
- Product
- DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy.
- Product code
- HRX Arthroscope
- Reason for recall
- Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
- FDA root cause
- Component change control
- Action
- DePuy Mitek inititaed telephone notification on Ausgust 29, 2013 and sent an Urgent Voluntary Product Recall letter dated September 5, 2013 to US and OUS affected customers/affiliates. The customers were notified of the affected products, problem and actions to be taken. The customers were instructed to immediately check all inventories to locate and return affected product following the enclosed instructions. Stericycle is cooridianting the returns. If you have any questions or concerns in regards to this recall, please contact Stericycle directly at 1-866-737-1928.
- Quantity in commerce
- 29280 units
- Distribution
- Worldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
- Product codes and lots
- All lot codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0191-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0192-2014 | Class II | Terminated | Voluntary: Firm initiated |