510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2285-2013Class II

SmartSite Low Sorbing Infusion Set, Model# 10015048. Intravascular administration set

CareFusion 303, Inc / initiated 2013-08-30 / Terminated / root cause: manufacturing process

CareFusion 303, Inc. began a recall of SmartSite Low Sorbing Infusion Set, Model# 10015048. Intravascular administration set on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model# 10015048, because there is a potential for separation at the tubing sleeve below the drip chamber". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2285-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
SmartSite Low Sorbing Infusion Set, Model# 10015048. Intravascular administration set.
Product code
FPA Set, Administration, Intravascular
Reason for recall
CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model# 10015048, because there is a potential for separation at the tubing sleeve below the drip chamber.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A customer letter dated August 30, 2013 will be sent to customers to inform them that CareFusion is recalling the SmartSite Low Sorbing Infusion Set Models 10015048, Lot Numbers 130153883 and 13025342 as a result of a potential for separation at the tubing sleeve below the drip chamber.The letter informs the customers of the problem identified, issues, potential risk, and the immediate actions to be taken. Customers are instructed to complete the customer response form and return the form to CareFusion. Customers are instructed to return recalled products directly to distributors. A distributor letter will also be sent to distributors on 8/30/13 to inform them of the problems identified and the actions to be taken. Distributors are instructed to contact the Customer Support Team, return customer response form, and they are instructed to contact the CareFusion Support Center at 1-800-562-6018 with any questions. .
Quantity in commerce
15860 units
Distribution
Nationwide in US: PA, OK, TX, FL, NC, CA, IN, WI
Product codes and lots
Lot Numbers 13015883 and 13025342.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2285-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2285-2013Class IITerminatedVoluntary: Firm initiated