ABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions...
Horiba Instruments, Inc dba Horiba Medical began a recall of ABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions: intracellularly in muscle contraction and glycogen... on . FDA classified it as Class III and gives the reason as "HORIBA Medical is informing all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2280-2013
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- ABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions: intracellularly in muscle contraction and glycogen metabolism, extracellularly, in bone mineralization, in blood coagulation and in transmission of nerve impulses. Calcium is present in plasma in three forms: free, bound to proteins or complexed with anions as phosphate, citrate and bicarbonate. Under physiological conditions, calcium balance is determined by the relationship between calcium intake and calcium absorption and excretion. Urinary excretion is an important determinant of calcium retention in the body. Decreased total calcium levels can be associated with diseases of the bone apparatus (especially osteoporosis), kidney diseases (especially under dialysis), defective intestinal absorption and hypoparathyroidism. Increased total calcium can be measured in hyperparathyroidism, malignant diseases with metastases and sarcoidosis. Calcium measurements also help in monitoring of calcium supplementation mainly in the prevention of osteoporosis.
- Product code
- CIC Cresolphthalein Complexone, Calcium
- Reason for recall
- HORIBA Medical is informing all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent.
- FDA root cause
- Under Investigation by firm
- Action
- Horiba Medical sent an Urgent Safety Notice dated September 6, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer notification letter dated September 6, 2013 will be sent to all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent (Reference #: A11A01633) Lot#: 014664503. The recall letter informs the customers that the initial functional tests indicate that there is no interference identified with regards to the patient results. However, with time, there may be an increase in the quantity of precipitates and this could potentially interfere with the performance of this reagent. Customers are informed of the actions to be taken and to complete the customer response form within 10 days. Further questions please call (949) 453-0500 Ext. 208
- Quantity in commerce
- 2,658 units ( 1240 US 1418 OUS )
- Distribution
- Worldwide Distribution - USA including the country of Canada.
- Product codes and lots
- Lot/Serial #(s): 014664503
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2280-2013 | Class III | Terminated | Voluntary: Firm initiated |