Z-2271-2013Class II

Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients...

Stryker Corp. / initiated 2013-08-30 / Terminated / root cause: design

Stryker Endoscopy began a recall of Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients. Model number 0240-095-200, manufactured by... on . FDA classified it as Class II and gives the reason as "When positioning the display located on top of the video cart, it is possible for the user to apply excessive force and push the display up and out of the mounting arm. This can result in the display falling upon the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2271-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients. Model number 0240-095-200, manufactured by Stryker Endoscopy San Jose The intended use of the Stryker Vision Mount is to hold and position an LCD monitor. The Stryker Vision Mounting Arm can be used with any Stryker flat panel display and is It is compatible with Stryker Carts (0240099001, 0240099011 or 0240099020). It is indicated for use in a surgical operating room environment and is exposed to environmental conditions associated with the storage and shipping of the product. It is not intended to enter the sterile field. The Stryker Vision Mount does not deliver to and/or extract from the patient any energy or material. It does not process any biological materials, as it is not intended to be sterilized by the user. The Stryker Vision Mount is only used to display surgical images and video and does not sustain or support life.
Product code
BZN Cart, Emergency, Cardiopulmonary (Excluding Equipment)
Reason for recall
When positioning the display located on top of the video cart, it is possible for the user to apply excessive force and push the display up and out of the mounting arm. This can result in the display falling upon the user/patient.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker Endiscopy sent an Urgent Device Correction letter dated August 30, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inform individuals within their organization who needed to be aware of this Medical Device Field Correction, contact Stericycle at 1-877-546-0482 or email [email protected] to schedule a time for Stericycle to perform the correction, or perform the correction themselves by requesting the materials and following the correction instructions included. Customers were also instructed to report all adverse events or product quality problems to Stryker Endoscopy at 1-800-624-4422, press 3 for Customer Service. --UPDATE-- On September 17, 2013, the recall was expanded to include additional customers who may have received the screw and washer included with the design change update. Notification was sent to these consignees on September 17, 2013. .
Quantity in commerce
14271 units
Distribution
Worldwide Distribution - USA including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and internationally to Australia, Canada, Chile, China, Argentina, Guam, Taiwan, India, Japan, Korea, Latin America, Philippines, Mexico, New Zealand, Hong Kong, Singapore, and Switzerland.
Product codes and lots
Model number 0240-095-200, all serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2271-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2271-2013Class IITerminatedVoluntary: Firm initiated