Z-0343-2014Class II

ADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and specialty chemistries...

Siemens Healthcare Diagnostics Inc. / initiated 2013-09-03 / Terminated / root cause: software

Siemens Healthcare Diagnostics began a recall of ADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0343-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA 1200 System Software: 1) V2.00, 2) V2.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems.
Root cause tag
software
FDA root cause
Software design
Action
On September 3, 2013, Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice for the ADVIA 1200, 1650, 1800 and 2400 Chemistry Systems due to a software issue. The software issue has caused the Chemistry Systems to generate reports without error flags. Notices were sent to domestic customers via U.S. and outside the US the distribution is determined at the country level. The customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue at 877-229-3711 (prompt 18).
Quantity in commerce
625 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Vietnam.
Product codes and lots
Product Codes: 1) 10386841, 2) 10469445

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0343-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0343-2014Class IITerminatedVoluntary: Firm initiated
Z-0344-2014Class IITerminatedVoluntary: Firm initiated
Z-0345-2014Class IITerminatedVoluntary: Firm initiated
Z-0346-2014Class IITerminatedVoluntary: Firm initiated