510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0374-2014Class II

(Siemens Rad Fluoro Uro Systems) Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software...

Siemens Medical Solutions USA, Inc. / initiated 2013-09-03 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of (Siemens Rad Fluoro Uro Systems) Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10 Image intensified fluoroscopic x-ray... on . FDA classified it as Class II and gives the reason as "A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0374-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
(Siemens Rad Fluoro Uro Systems) Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10 Image intensified fluoroscopic x-ray system
Product code
KPR System, X-Ray, Stationary
Reason for recall
A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Healthcare sent a Field Safety Notice dated September 3, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Field Safety Notice to any new users and provide the new owner information to Siemens Healthcare. Customers were asked to send back the attached form confirming receipt of the Field Safety Notice. Customers with questions should call 610-850-4634. For questions regarding this recall call 610-219-6300.
Quantity in commerce
357
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Canada.
Product codes and lots
Model numbers: 10094200, 10094910, 10281013, 10281163, 10502200

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0374-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0374-2014Class IITerminatedVoluntary: Firm initiated