ADVIA 2400 System Software V4.01. Performs assays for general and specialty chemistries, methods for specific...
Siemens Healthcare Diagnostics began a recall of ADVIA 2400 System Software V4.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0346-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA 2400 System Software V4.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
- Product code
- JGS Electrode, Ion Specific, Sodium
- Reason for recall
- Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On September 3, 2013, Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice for the ADVIA 1200, 1650, 1800 and 2400 Chemistry Systems due to a software issue. The software issue has caused the Chemistry Systems to generate reports without error flags. Notices were sent to domestic customers via U.S. and outside the US the distribution is determined at the country level. The customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue at 877-229-3711 (prompt 18).
- Quantity in commerce
- 876 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Vietnam.
- Product codes and lots
- Product Code: 10639265
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0343-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0344-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0345-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0346-2014 | Class II | Terminated | Voluntary: Firm initiated |