Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral...
Fujinon Inc began a recall of Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for... on . FDA classified it as Class II and gives the reason as "A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0112-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis and endoscopic treatment.
- Product code
- FDS Gastroscope And Accessories, Flexible/Rigid
- Reason for recall
- A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope.
- Root cause tag
- manufacturing process
- FDA root cause
- Unknown/Undetermined by firm
- Action
- The firm, FUJIFILM Medical Systems, USA, Inc. (FMSU-ESD) sent an "URGENT-MEDICAL DEVICE CORRECTION (UMDC)" and Customer Acknowledgement forms dated July 31, 2013 to their customers. The letter described the issue with the affected product; instructed customers on how to identify affected units; instructed customers to immediately cease use of the affected units; return affected units to FMSU-ESD for repairs and quality control inspection; and complete and return the attached "Customers Acknowledgement" form to FMSU-ESD, Attn: Regulatory Compliance, via fax to (973) 686-2616 or email: [email protected]. If you have any questions regarding this notice, please do not hesitate to contact Director, Quality and Regulatory Compliance, at (973) 686-2636 or by email: [email protected].
- Quantity in commerce
- 12 Units
- Distribution
- Worldwide distribution: US (nationwide) including states of: NJ, PA and VA; and country of Canada.
- Product codes and lots
- Model Number EG-530UT2. Serial Numbers: NU038A018, NU038A019, NU038A022, NU038A027, NU038A028 and NU038A029
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0112-2014 | Class II | Terminated | Voluntary: Firm initiated |