chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the...
bioMerieux, Inc. began a recall of chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens on . FDA classified it as Class II and gives the reason as "Complaints for Salmonella strains growing on this medium producing an uncharacteristic white colored colony instead of the expected pale pink to mauve colored colonies. This could potentially lead to Salmonella colonies...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0087-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens.
- Reason for recall
- Complaints for Salmonella strains growing on this medium producing an uncharacteristic white colored colony instead of the expected pale pink to mauve colored colonies. This could potentially lead to Salmonella colonies not being detected.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Recall letters were issued to the US customers by Biomeriux, Durham, NC on 9/6/2013.
- Quantity in commerce
- 90
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot 1002234210 and 1002300040