Z-0776-2014Class II
Arena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of the...
SPINEFRONTIER, Inc. dba LESspine Innovations / initiated 2013-09-24 / Terminated
SpineFrontier, Inc. began a recall of Arena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of the implant, manufactured from 17-4 Stainless Steel into... on . FDA classified it as Class II and gives the reason as "Handles and heads may not be compatible". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0776-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SPINEFRONTIER, Inc. dba LESspine Innovations
- Product
- Arena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of the implant, manufactured from 17-4 Stainless Steel into 35X26mm (11-81004-XX) and 38X28mm (11-81012-XX) varieties, each with multiple height configurations. The Trial Heads are intended to attach tot he Trial Handle (11-81002) and to be inserted into the intervertebral disc space. The Trial Handle consists of a knurled handle and cylindrical shaft with an M4 threads interface at the distal end; 17-4 Stainless steel is used as the material for the handles. The Trial Heads distract the disc space to the surgeon's desired height to gauge which size implant should be utilized to complete the Anterior Lumbar Interbody Fusion Procedure.
- Product code
- HTE Elevator
- Reason for recall
- Handles and heads may not be compatible.
- FDA root cause
- Other
- Action
- The firm, SpineFrontier, sent an "URGENT Arena-L Trials Voluntary Recall Notification" letter dated September 24, 2013 sent to its consignee/customer. The letter described the product, problem and actions to be taken. The consignee/customer were instructed to confirm receipt of the notice and contact Customer Service to initiate the immediate return of the listed parts. In addition, consignee/customer were instructed to complete a recall ACKNOWLEDGEMENT FORM and return it to SpineFrontier via fax to: Attention: Customer Service at 978-232-3991 or email to: [email protected]. Please feel free to contact us at 978-232-3990 Monday through Friday 8:30AM to 5:30PM EST with any questions.
- Quantity in commerce
- 23 devices in total
- Distribution
- US Distribution: VA only.
- Product codes and lots
- CT128047, CT128048
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0776-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0777-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0778-2014 | Class II | Terminated | Voluntary: Firm initiated |