Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0048-2014Class II

STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is...

Advanced Medical Optics, Inc. / initiated 2013-09-16 / Terminated / root cause: manufacturing process

Abbott Medical Optics, Inc. began a recall of STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing... on . FDA classified it as Class II and gives the reason as "This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0048-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
STAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
Product code
LZS Excimer Laser System
Reason for recall
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Abbott Medical Optics, Inc., (AMO), sent an "ADVISORY NOTICE and a COVER LETTER FOR ADVISORY NOTICE" dated September 16, 2013 to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the Star Notification Acknowledgement form via fax to 1-714-247-4530 within 3 business days or mail to Abbott Medical Optics, 510 Cottonwood Drive, MS 347118, Milpitas, CA 95035, and share this advisory notice with your staff. If you have any questions related to operation of STAR Excimer Laser System or documentation requirements associated with this letter, please call AMO Technical Support at 1-800-511-0911 and you will be directed to an AMO Representative.
Quantity in commerce
834 units in total
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.
Product codes and lots
All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0048-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0046-2014Class IITerminatedVoluntary: Firm initiated
Z-0047-2014Class IITerminatedVoluntary: Firm initiated
Z-0048-2014Class IITerminatedVoluntary: Firm initiated
Z-0049-2014Class IITerminatedVoluntary: Firm initiated