Z-0091-2014Class I

Edwards Lifesciences EMBOL-X Glide Protection System; Model Numbers: EXGF24D, EXGF24LLD, EXGF24MMD...

Edwards Lifesciences LLC / initiated 2013-09-16 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of Edwards Lifesciences EMBOL-X Glide Protection System; Model Numbers: EXGF24D, EXGF24LLD, EXGF24MMD, EXGF24SSD, EXGF24XLD, EXGF24XSD, EXGFXS2D, EXGF24SS2D... on . FDA classified it as Class I and gives the reason as "Edwards Lifesciences is recalling certain lots of EMBOL-X Glide Protection System due to deformation of the cannula tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0091-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards Lifesciences EMBOL-X Glide Protection System; Model Numbers: EXGF24D, EXGF24LLD, EXGF24MMD, EXGF24SSD, EXGF24XLD, EXGF24XSD, EXGFXS2D, EXGF24SS2D, EXGF24MM2D, EXGF24LL2D and EXGF24XL2D. The EMBOL-X access device is indicated for the perfusion of the ascending aorta during short-term (less than or equal to0 6 hours) cardiopulmonary bypass (CPB) surgery where procedures may require the hemostatic introduction and removal of compatible intravascular devices into the vascular system.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Edwards Lifesciences is recalling certain lots of EMBOL-X Glide Protection System due to deformation of the cannula tip.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Edwards Lifesciences, sent an "URGENT FIELD SAFETY NOTICE PRODUCT RECALL-ACTION REQUIRED" letter dated September 16, 2013 via Fed-Ex to its consignees/customers. Consignees were asked to review inventory for the cannulae and remove the cannulae from any kits received; quarantine affected material at site and return product to Edwards LIfesciences, Attn: Santosh Bhagat, 12050 Lone Peak Drive, Draper, UT 84020, Attention: RECALL, RGA #; call Edwards Customer Service at 800-424-3278 to obtain RGA# and replacement product (note: the filter is not affected and does not have to be returned with the cannulae); and complete and return the attached acknowledgment form via fax to Edwards Customer Service at 800-422-9329 within three days of receipt of this Field Safety Notice. If you have any questions, please call Edwards Customer Service at 800-424-3278 from the hours 6:00am - 4:30pm PST or contact your Edwards' sales representative concerning the recall.
Quantity in commerce
12,034
Distribution
Worldwide distribution: US (nationwide) and country of: Canada.
Product codes and lots
Model No: EXGF24D, EXGF24LLD, EXGF24MMD, EXGF24SSD, EXGF24XLD, EXGF24XSD, EXGFXS2D, EXGF24SS2D, EXGF24MM2D, EXGF24LL2D and EXGF24XL2D.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0091-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0091-2014Class ITerminatedVoluntary: Firm initiated