Z-0033-2014Class II

Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T...

Philips North America LLC / initiated 2013-08-07 / Terminated / root cause: design

Philips Medical Systems (Cleveland) Inc began a recall of Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for... on . FDA classified it as Class II and gives the reason as "There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0033-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system.
Root cause tag
design
FDA root cause
Device Design
Action
Philips Healthcare sent an Urgent Medical Device Correction letter via certified mail on August 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information with all of their staff who need to be aware of the contents of this communication. Philips Healthcare will perform any necessary corrective actions prior to the customer's next cryogen refill. Customers with questions were instructed to contact their Philips field service representative: Technical Support Line 1-800-722-9377. For questions regarding this recall call 1-800-722-9377.
Quantity in commerce
Total 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panorama 0.6T: 55 units)
Distribution
Worldwide Distribution - US (nationwide) including FL, CO, KY, MS, PA, NC, NJ, TX and Puerto Rico and internationally to Italy.
Product codes and lots
M/N: 781220, 781230, 781301, 78129 and 78127. Codes: AM128, 1510, 1518, BA318, AM120, ZA178, ZA196, ZA107, ZA133, ZA172, ZA158, BA149 and ZA208.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0033-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0033-2014Class IITerminatedVoluntary: Firm initiated