Z-1845-2014Class II

Siemens RAD Fluoro Uro System (Ysio, Luminos dRF, and Uroskop Omnia) with software versions VB10G

Siemens Medical Solutions USA, Inc. / initiated 2013-08-28 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens RAD Fluoro Uro System (Ysio, Luminos dRF, and Uroskop Omnia) with software versions VB10G. Product Usage: Image Intensified Fluoroscopic X-ray system on . FDA classified it as Class II and gives the reason as "When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1845-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens RAD Fluoro Uro System (Ysio, Luminos dRF, and Uroskop Omnia) with software versions VB10G. Product Usage: Image Intensified Fluoroscopic X-ray system
Product code
KPR System, X-Ray, Stationary
Reason for recall
When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a notification letter, dated August 28, 2013, to Affected customers. The letter identified the affected product, problem and actions to be taken. The letter provided work-around instructions and announced that a field modification was prepared to resolve the issue.
Quantity in commerce
76
Distribution
USA Nationwide Distribution in the states of OH, NJ, MA, CO, NY, CA, MD, OR, KY, PA, IL, VA, TX, MN, KS, GA, and WY.
Product codes and lots
Model numbers 10094910, 10281013, and 10281163 with software version VB10G.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1845-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1845-2014Class IITerminatedVoluntary: Firm initiated