Z-0657-2014Class II

***REF IMRI-MOTOR***iMRI Handpiece. Use with iMRI System Only***Rx Only***Manufacturer: The Anspach Effort...

The Anspach Effort, Inc. / initiated 2013-04-25 / Terminated / root cause: design

The Anspach Effort, Inc. began a recall of ***REF IMRI-MOTOR***iMRI Handpiece. Use with iMRI System Only***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone on . FDA classified it as Class II and gives the reason as "The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0657-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
***REF IMRI-MOTOR***iMRI Handpiece. Use with iMRI System Only***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.
Product code
HBB Motor, Drill, Pneumatic
Reason for recall
The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, Anspach sent an "URGENT" Medical Devics Removal" letter dated April 24, 2013, to its customers. Customer Immediate Actions: 1. Screen your current inventory and return all Anspach Pneumatic Drill System without a pressure Relief valve (PRV) from your inventory. 2. If you have any product listed on the attached form to arrange for immediate return and replacements. 3. complete the attached Customer Reply form indicating the product being returned. 4. Upon receipt of a replacement product, equipment being returned must be received by Anspach within 10 business days of the replacement request. Should you have any queries, please contact Anspach Product Support at (800) 327-6887.
Quantity in commerce
12
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and states of: AL, AR, AZ, CA, CO, CT, DC, DE, GA, FL, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VE, VT, WA, WI, WY, and WV. Internationally to: AT, AU, BR, BS, CA, CH, CL, CO, CZ, CR, DE, DK, DO, EG, ES, FI, FR, GB, GR, HK, HU, ID, IL, IN, IS, JP, KH, KR, LB, LT, MX, MY, NL, NO, NZ, PA, PE, PH, PL, PT, SA, SG, SY, TH, TR, TW, and ZA.
Product codes and lots
Serial Number F37306977201

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0657-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0654-2014Class IITerminatedVoluntary: Firm initiated
Z-0655-2014Class IITerminatedVoluntary: Firm initiated
Z-0656-2014Class IITerminatedVoluntary: Firm initiated
Z-0657-2014Class IITerminatedVoluntary: Firm initiated