Z-0178-2014Class II

Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit...

VitalScientific / initiated 2013-09-04 / Terminated / root cause: design

SEPPIM SAS began a recall of Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL... on . FDA classified it as Class II and gives the reason as "Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0178-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
VitalScientific
Product
Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL. The vial's label shows CALCIUM ARSENAZO, CALA-5220, 20 mL, . CALCIUM ARSENAZO is intended use for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). It is not intended for use in Point of Care setting.
Product code
CJY Azo Dye, Calcium
Reason for recall
Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum.
Root cause tag
design
FDA root cause
Device Design
Action
ELITech Group of Bothell, WA sent an Urgent: Product Recall Notice dated September 9, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: " Laboratories must not use plasma for the determination of total calcium when using this product. The product can still be used to determine total calcium with serum sample. " ELITech Clinical Systems SAS has revised the package insert and kit box label for the CALCIUM ARSENAZO reagent reference CALA-0250. All mentions of the use of the plasma will be removed from the documentation. If you have inventory of CALCIUM ARSENAZO, reference CALA- 0250. All lots. in your laboratory, please discard the insert that you have and replace it with the insert enclosed with this notice. " If you have transferred any of this product to another location, please forward a copy this bulletin to that location. " All users receive CALCIUM ARSENAZO, CALA-0250, from distributors. To be certain everyone who is possibly affected by this change is contacted, we are sending this communication to all registered users. As confirmation of receipt of this notice, please return the attached receipt confirmation to Technical Support by fax (401-642-9003) within 48 hours of receipt. We apologize for the inconvenience that this situation can cause in your laboratories; we remain at your disposal for further information. If you have any questions, please contact Technical Support at 1-855-ELITECH (855-354-324) We thank you for your cooperation and the attention which you will give to this bulletin.
Quantity in commerce
47 kits in US and 48 kits in Sweden
Distribution
Worldwide Distribution - US including the states of KY and UT and the country of Sweden.
Product codes and lots
Lot numbers distributed in the US:  Lot 11-2614, Expiration 07/2013. Lot 12-1002; Expiration 03/2014. Lot 12-2220, Expiration 08/2014.  Lot numbers distributed in Sweden: 11-2129, 12-2616, 11-2769, 12-2975, 10-3036, 10-0454, 10-1874, 11-0176, 11-1248, 11-1690, 12-0140, 12-0938, 12-1383, 13-0004, and 13-0408

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0178-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0178-2014Class IITerminatedVoluntary: Firm initiated