GENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO...
Smith & Nephew Inc began a recall of GENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic on . FDA classified it as Class II and gives the reason as "The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0444-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- GENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Smith & Nephew sent an Urgent Recall Notification letter dated August 27, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. They were instructed to cease use and distribution of the product and quarantine for return. Customers were instructed to contact Smith & Nephew via email at [email protected] or fax 901-566-7975 to obtain a return authorization (RA) number. Customers with questions were also instructed use this email address. Customers with affected product were instructed to return it along with the completed inventory Return Certification Form to the address listed on the form. Non-responding consignees were notified again on September 27, 2013, by email. For questions regarding this recall call 901-396-2121.
- Quantity in commerce
- 16 pieces
- Distribution
- Worldwide Distribution - USA including Nevada and Internationally to United Arab Emirates, Italy, Spain, Belgium, Bulgaria, Poland, France, Japan, Netherlands, Australia, Turkey, Russian Federation, Germany and United Kingdom.
- Product codes and lots
- Batch No. 12GT20908
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0444-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0445-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0446-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0447-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0448-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0449-2014 | Class II | Terminated | Voluntary: Firm initiated |