Z-0448-2014Class II

GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith &...

Smith & Nephew, Inc. / initiated 2013-08-27 / Terminated / root cause: packaging

Smith & Nephew Inc began a recall of GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic on . FDA classified it as Class II and gives the reason as "The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0448-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Smith & Nephew sent an Urgent Recall Notification letter dated August 27, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. They were instructed to cease use and distribution of the product and quarantine for return. Customers were instructed to contact Smith & Nephew via email at [email protected] or fax 901-566-7975 to obtain a return authorization (RA) number. Customers with questions were also instructed use this email address. Customers with affected product were instructed to return it along with the completed inventory Return Certification Form to the address listed on the form. Non-responding consignees were notified again on September 27, 2013, by email. For questions regarding this recall call 901-396-2121.
Quantity in commerce
46 pieces
Distribution
Worldwide Distribution - USA including Nevada and Internationally to United Arab Emirates, Italy, Spain, Belgium, Bulgaria, Poland, France, Japan, Netherlands, Australia, Turkey, Russian Federation, Germany and United Kingdom.
Product codes and lots
Batch No. 12GT20775

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0448-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0444-2014Class IITerminatedVoluntary: Firm initiated
Z-0445-2014Class IITerminatedVoluntary: Firm initiated
Z-0446-2014Class IITerminatedVoluntary: Firm initiated
Z-0447-2014Class IITerminatedVoluntary: Firm initiated
Z-0448-2014Class IITerminatedVoluntary: Firm initiated
Z-0449-2014Class IITerminatedVoluntary: Firm initiated