Z-2302-2013Class II

NC Sprinter; Rapid Exchange Balloon Dilatation Catheter; Sterile; 2.25 mm; Model NCSP22512X. Manufacturer...

Medtronic, Inc. / initiated 2013-09-16 / Terminated

Medtronic Vascular, Inc. began a recall of NC Sprinter; Rapid Exchange Balloon Dilatation Catheter; Sterile; 2.25 mm; Model NCSP22512X. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis... on . FDA classified it as Class II and gives the reason as "Compliance chart included in lot 207002011 of the NC Sprinter RX 2.25 x 12 mm products lists the size of the device as 2.5mm instead of the correct 2.25mm size". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2302-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
NC Sprinter; Rapid Exchange Balloon Dilatation Catheter; Sterile; 2.25 mm; Model NCSP22512X. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 The NC Sprinter Rapid Exchange Balloon Dilatation Catheter is a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System. The balloon at the distal end of the catheter can be inflated to a defined diameter at a specific pressure (see labelling). The proximal end of the catheter has a female luer for attachment to an inflation device. The catheter provides a lumen which enables the use of a guide wire to position the catheter. Radiopaque balloon marker(s) enable accurate placement. Shaft markers for brachial and femoral techniques are in place.
Product code
LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous
Reason for recall
Compliance chart included in lot 207002011 of the NC Sprinter RX 2.25 x 12 mm products lists the size of the device as 2.5mm instead of the correct 2.25mm size.
FDA root cause
Mixed-up of materials/components
Action
Medtronic sent an Urgent Medical Device Recall letter beginning September 16, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate that your facility has received affected NC Sprinter RX Balloon Dilatation Catheters as noted in the attached customer notification detail report. Consequently, Medtronic is asking you to take the following actions: 1. Remove and quarantine all affected catheters that remain in your inventory. 2. Return the affected catheters to Medtronic. A Medtronic representative can assist in facilitating the return and replacement of this product as necessary. 3. Complete the attached Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. Medtronic has taken the necessary steps to prevent any future shipment of mislabeled product. Regulatory agencies will be notified about this recall as applicable. Please share this notification with others in your organization as appropriate. If any NC Sprinter RX Balloon Dilatation Catheters within the scope of this recall have been forwarded to another facility, please notify that facility accordingly and facilitate the retrieval of the affected product. We appreciate your cooperation and apologize for the inconvenience that this issue may cause. Should you have any questions, please contact your local Medtronic representative.
Quantity in commerce
122
Distribution
Worldwide Distribution - USA and the countries of Australia, Belgium, Canada, Hong Kong, Malaysia, New Zealand, Singapore, Spain, Switzerland, Taiwan and Thailand.
Product codes and lots
Lot number 207002011; Exp 5-13-2015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2302-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2302-2013Class IITerminatedVoluntary: Firm initiated