Z-0148-2014Class II

Fresenius Liberty Cycler Part Number: RTLR180111

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2013-10-14 / Terminated / root cause: design

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal... on . FDA classified it as Class II and gives the reason as "Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0148-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis
Root cause tag
design
FDA root cause
Device Design
Action
Fresenius Medical Care contacted all affected consignees by telephone on 10/14/13 to inform them of the product safety alert. Follow-up Urgent Safety Alert letter was sent on 10/18/13 and 10/23/13 by Certified Mail, Return Receipt requested. The letter identified the affected product, problem and actions to be taken. Customers were instructed to report any events associated with this safety alert to Customer Service Team at 1-800-323-5188.
Quantity in commerce
16,600
Distribution
USA Nationwide Distribution
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0148-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0148-2014Class IITerminatedVoluntary: Firm initiated