Z-0541-2014Class II

Siemens Dimension Vista Homocysteine Flex Reagent Cartridge The HCYS method is an in vitro diagnostic test...

Siemens Healthcare Diagnostics Inc. / initiated 2013-09-18 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension Vista Homocysteine Flex Reagent Cartridge The HCYS method is an in vitro diagnostic test for the quantitative measurement of total... on . FDA classified it as Class II and gives the reason as "Dimension Vista HCYS lot 13064MA shows erratic recovery of QC and patient samples". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0541-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Dimension Vista Homocysteine Flex Reagent Cartridge The HCYS method is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and ETDA plasma on the Dimension Vista System. Measurements of homocysteine aid in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Product code
LPS Urinary Homocystine (Nonquantitative) Test System
Reason for recall
Dimension Vista HCYS lot 13064MA shows erratic recovery of QC and patient samples.
FDA root cause
Unknown/Undetermined by firm
Action
Siemens sent a Urgent Medical Device Recall letter dated September 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Actions to be taken by Customer: Please discontinue use and discard any remaining inventory of Dimension Vista HCYS Flex reagent cartridge (Catalog# K7044) lot 13064MA. Please complete the attached form and fax it to (302) 631-8467 to facilitate your no charge replacement needs. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation has caused. If you have any questions, please contact your Siemens Customer Care Center- Technical Solutions or your local Siemens technical support representative. For further questions please call (800) 441-9250.
Quantity in commerce
302
Distribution
Worldwide Distribution - US Distribution including the states of OH, MD, CO, NJ, IA, NC, VA, WI and DC., and the countries of Austria, Denmark, France, Italy, Netherlands and Norway.
Product codes and lots
lot 13064MA, expiration 2014-03-05

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0541-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0541-2014Class IITerminatedVoluntary: Firm initiated