Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0316-2014Class II

VITROS Chemistry Products Li Slides packaged as 300 slides/pack catalog number 8318925 and packaged as 90...

Ortho-Clinical Diagnostics, Inc. / initiated 2013-06-12 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products Li Slides packaged as 300 slides/pack catalog number 8318925 and packaged as 90 slides/pack catalog number 1632660 VITROS Li Slides... on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0316-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Li Slides packaged as 300 slides/pack catalog number 8318925 and packaged as 90 slides/pack catalog number 1632660 VITROS Li Slides quantitatively measure lithium (Li) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Product code
JIH Flame Photometry, Lithium
Reason for recall
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.
Quantity in commerce
62,039 units (5,974 units domestically & 4,702 units internationally)
Distribution
Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.
Product codes and lots
Multiple lots affected, with expiry dates through 22-Mar-2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0316-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0292-2014Class IITerminatedVoluntary: Firm initiated
Z-0293-2014Class IITerminatedVoluntary: Firm initiated
Z-0294-2014Class IITerminatedVoluntary: Firm initiated
Z-0295-2014Class IITerminatedVoluntary: Firm initiated
Z-0296-2014Class IITerminatedVoluntary: Firm initiated
Z-0297-2014Class IITerminatedVoluntary: Firm initiated
Z-0298-2014Class IITerminatedVoluntary: Firm initiated
Z-0299-2014Class IITerminatedVoluntary: Firm initiated
Z-0300-2014Class IITerminatedVoluntary: Firm initiated
Z-0301-2014Class IITerminatedVoluntary: Firm initiated
Z-0302-2014Class IITerminatedVoluntary: Firm initiated
Z-0303-2014Class IITerminatedVoluntary: Firm initiated
Z-0304-2014Class IITerminatedVoluntary: Firm initiated
Z-0305-2014Class IITerminatedVoluntary: Firm initiated
Z-0306-2014Class IITerminatedVoluntary: Firm initiated
Z-0307-2014Class IITerminatedVoluntary: Firm initiated
Z-0308-2014Class IITerminatedVoluntary: Firm initiated
Z-0309-2014Class IITerminatedVoluntary: Firm initiated
Z-0310-2014Class IITerminatedVoluntary: Firm initiated
Z-0311-2014Class IITerminatedVoluntary: Firm initiated
Z-0312-2014Class IITerminatedVoluntary: Firm initiated
Z-0313-2014Class IITerminatedVoluntary: Firm initiated
Z-0314-2014Class IITerminatedVoluntary: Firm initiated
Z-0315-2014Class IITerminatedVoluntary: Firm initiated
Z-0316-2014Class IITerminatedVoluntary: Firm initiated
Z-0317-2014Class IITerminatedVoluntary: Firm initiated
Z-0318-2014Class IITerminatedVoluntary: Firm initiated
Z-0319-2014Class IITerminatedVoluntary: Firm initiated
Z-0320-2014Class IITerminatedVoluntary: Firm initiated
Z-0321-2014Class IITerminatedVoluntary: Firm initiated
Z-0322-2014Class IITerminatedVoluntary: Firm initiated
Z-0323-2014Class IITerminatedVoluntary: Firm initiated
Z-0324-2014Class IITerminatedVoluntary: Firm initiated
Z-0325-2014Class IITerminatedVoluntary: Firm initiated
Z-0326-2014Class IITerminatedVoluntary: Firm initiated
Z-0327-2014Class IITerminatedVoluntary: Firm initiated
Z-0328-2014Class IITerminatedVoluntary: Firm initiated