Z-0291-2014Class II

Access 2 Immunoassay System, Part Number: 81600N The Access Immunoassay System is an in vitro diagnostic...

Siemens Healthcare Diagnostics Inc. / initiated 2013-09-12 / Terminated

Beckman Coulter Inc. began a recall of Access 2 Immunoassay System, Part Number: 81600N The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is recalling the Access 2 Immunoassay System because they determined that the fuses in the system may be of the wrong amperage". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0291-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access 2 Immunoassay System, Part Number: 81600N The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter is recalling the Access 2 Immunoassay System because they determined that the fuses in the system may be of the wrong amperage.
FDA root cause
Mixed-up of materials/components
Action
A phone call was made on 9/12/13 to customer who received the Access 2 System. The phone script explains the problem identified of a possible wrong fuse and may result in failure of the fuse to operate properly. Customer was instructed that their Field Service Engineer, will call them back the following week to check on their system and replace the fuses if necessary. If you have any questions concerning this information, please call 952-393-8546.
Quantity in commerce
8 units total (1 unit in the US)
Distribution
Worldwide distribution: US (nationwide) in the state of: TX and countries of: Colombia, Hong Kong and Korea.
Product codes and lots
Serial Numbers:  508801, 508802, 508803, 508804, 508805, 508806, 508807, 508810

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0291-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0291-2014Class IITerminatedVoluntary: Firm initiated