Synthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN...
Synthes USA HQ, Inc. began a recall of Synthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN Nail is intended to treat stable and unstable... on . FDA classified it as Class II and gives the reason as "Certain lots of Trochanteric Fixation Nail were distributed to the field with the incorrect labeling of the length and diameter of the nail on the packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0474-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN Nail is intended to treat stable and unstable fractures of the proximal femur.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Certain lots of Trochanteric Fixation Nail were distributed to the field with the incorrect labeling of the length and diameter of the nail on the packaging.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following actions: Call Syntehs at 1-800-479-6329 to obtain a Return Authorization Number. Complete the Verification Section checking the appropriate box indicating affected product has been located. Return the Verfication Form with the product to: Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132. If no affected product found customers were instructed to take the following steps: Complete the attached Verification Section at the end of the letter by checking the appropriate box indicating no affected product has been located. Return the documents to Synthes by fax to 610-251-9005 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
- Quantity in commerce
- 12
- Distribution
- Nationwide Distribution including WA, FL, LA, TX, GA, IN, OH, CA, MI, and PA.
- Product codes and lots
- Part 456.475S, lot # 7121831 and Part 456.637S, lot #7121832, expiration 11/30/21.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0474-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY