Z-0408-2014Class II

Slipped Capital Femoral Epiphysis System

Synthes GmbH / initiated 2013-07-31 / Terminated / root cause: design

Synthes USA HQ, Inc. began a recall of Slipped Capital Femoral Epiphysis System Product Usage: Intended for fracture fixation of large bones and large bone fragments and for slipped capital femoral... on . FDA classified it as Class II and gives the reason as "There is a possibility for the Guide Wires of the Slipped Capital Femoral Epiphysis System to stick in the cannulation of the drill bit and the drill bits to break during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0408-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Slipped Capital Femoral Epiphysis System Product Usage: Intended for fracture fixation of large bones and large bone fragments and for slipped capital femoral epiphysis, pediatric femoral neck fractures, intercondylar femur fractures and sacroiliac joint disruptions.
Product code
HTW Bit, Drill
Reason for recall
There is a possibility for the Guide Wires of the Slipped Capital Femoral Epiphysis System to stick in the cannulation of the drill bit and the drill bits to break during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their inventory, immediately remove the affected product and return to Credit/Returns, Synthes. For questions call 610-719-5450.
Quantity in commerce
166
Distribution
USA Nationwide Distribution
Product codes and lots
Parts 03.207.001 and 03.207.008 with lot numbers:   Part # Lot # 03.207.001 PE00376  PE00439  PE01391  PE01329  PE01412  PE00504 03.207.008 PE00440  PE00377  PE00513  PE01475  PE01691

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0408-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0408-2014Class IITerminatedVoluntary: Firm initiated