Z-0452-2014Class II
Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the...
Synthes GmbH / initiated 2013-07-30 / Terminated / root cause: manufacturing process
Synthes USA HQ, Inc. began a recall of Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur on . FDA classified it as Class II and gives the reason as "Two lots of Titanium Trochanteric Fixation Nail (TFN) System are missing the second bend in the nail". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0452-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Two lots of Titanium Trochanteric Fixation Nail (TFN) System are missing the second bend in the nail.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 if they have any of the affeccted product to obtain a Return Authorization Number, Complete the Verification Section of the letter by checking the appropriate box indicating the affected product has been located. Return the Verification Form with the product to: Credit/Returns Synthes 1101 Synthes Avenue Monument, CO 80132. Customers should complete and return the Verification Frorm even if they do not have any of the affected product and return by fax to 1-888-386-2077 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450 or contact their Synthes Trauma Sales Consultant. For questions regarding this recall call 610-719-5000.
- Quantity in commerce
- 17
- Distribution
- Nationwide Distribution including IL, CA, AZ, FL, TX, UT, VA, NY, NM, and MT.
- Product codes and lots
- Part # 456.391S with lot numbers 5536930, exp. 5/20/2016, and 7161401, exp. 12/31/2021.