Z-0457-2014Class II

Product is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product of...

Abbott Medical / initiated 2013-07-17 / Terminated / root cause: manufacturing process

Hospira Inc. began a recall of Product is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product of China; Hospira, Inc.,; Lake Forest, IL 60045 USA... on . FDA classified it as Class II and gives the reason as "Hospira has recieved reports that the vacuum needed to create suction with the 1L Receptal System cannot be achieved due to the canister and liner not properly seating". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0457-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Product is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product of China; Hospira, Inc.,; Lake Forest, IL 60045 USA. Used to isolate suction waste during surgeries, or to clear patient airways, throat or wounds to allow treatment.
Product code
GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Reason for recall
Hospira has recieved reports that the vacuum needed to create suction with the 1L Receptal System cannot be achieved due to the canister and liner not properly seating.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An URGENT DEVICE RECALL letter dated July 17, 2013 was mailed to direct consignees. The letter requested the consignees: 1) quarantine any affected products, 2) complete and return the reply form included with the letter, 3) notify healthcare professionals within the consignee's organization of the recall, 4) return the affected products to Stericycle in accordance with the instructions in the letter. Consignees are further requested to forward a copy of the notification materials to any customers that they may have distributed the products to.
Quantity in commerce
3,560 units distributed from June 2004 - December 2012
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, FL, GA, IL, IN, KS, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, PA, SD, TX, UT, VA, and WA and countries of: Canada and EMEA.
Product codes and lots
*** DOMESTIC PRODUCT ***   List Numbers: 43449-04-11 and 43449-04-01;  Lot Numbers: All;   *** FOREIGN PRODUCT ***   List Numbers: 0L21297W1, 0L2129701, 0L2139701, 43449001L, 434490101, 434490102, 434490181, 434490181, 434490182, 4344901DX, 434496101, 434496161, 4344961A1, 434499090;  Lots: All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0457-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0457-2014Class IITerminatedVoluntary: Firm initiated