Z-0210-2014Class II

Composite Series Skull Clamp, one clamp per tote. The MAYFIELD Skull Clamp (A3059) is a cranial stabilization...

Integra LifeSciences Corporation / initiated 2013-08-29 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of Composite Series Skull Clamp, one clamp per tote. The MAYFIELD Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal... on . FDA classified it as Class II and gives the reason as "An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0210-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Composite Series Skull Clamp, one clamp per tote. The MAYFIELD Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary
Product code
HBL Holder, Head, Neurosurgical (Skull Clamp)
Reason for recall
An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.
Root cause tag
design
FDA root cause
Device Design
Action
Integra sent an Urgent: Voluntary Medical Device Recall Notification dated August 29, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Your local Integra NeuroSpecialist will set up a time with you to discuss replacement options but in the interim, Integra is asking that you do the following: 1. Review your inventory and determine if you have any MAYFIELD Composite Series Skull Clamps (A3059). If so, stop using them immediately and remove them from service to prevent any inadvertent use of the Composite Series Skull Clamp. Either while the Integra NeuroSpecialist is on-site or independently; 2. Complete the attached form. If you do not have affected product, check the box: I do not have Composite Series Skull Clamps. 3. If you do have affected product, check the box: I do have Composite Series Skull Clamps. Record the total quantity and Serial No.(s) of all Skull Clamps in your inventory. The Serial No. is located on the clamp upright comprised of two letters (SS) followed by six numerals. 4. Complete the other information and return by email or fax as indicated on the form, or give it to the Integra NeuroSpecialist who will return it for you. For assistance returning and exchanging these products or any other questions that you may have, please contact Integra Customer Service at 1-855-532-1723. We sincerely apologize for any inconvenience this may have caused and thank you for your cooperation.
Quantity in commerce
37 units
Distribution
US Distribution including the states of AL, AR, CA, CT, FL, GA, IA, IL, MS, NY, OH, SC, TN, and TX.
Product codes and lots
The following serial numbers are affected by this recall: SS120076, SS120082, SS120083, SS120084, SS120088, SS120089, SS120090, SS120092, SS120093, SS120094, SS120095, SS120111, SS120112, SS120113, SS120114, SS120115, SS120116, SS120117, SS120118, SS120119, SS120120, SS120121, SS120122, SS120123, SS120124, SS120125, SS120126, SS120127, SS120128, SS120130, SS120131, SS120133, SS120134, SS120135, SS120171, SS120252, SS120256, SS120257

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0210-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0210-2014Class IITerminatedVoluntary: Firm initiated