Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that...
Stryker Instruments Div. of Stryker Corporation began a recall of Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an... on . FDA classified it as Class II and gives the reason as "Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0127-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.
- Product code
- HRX Arthroscope
- Reason for recall
- Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 11/4/13, an URGENT MEDICAL DEVICE Recall Notification was sent to all domestic consignees via mail and foreign by email. An advisory was sent to every physician that uses any of the products listed. Every physician that uses any of these products must print their name and sign below which means you acknowledge that youve received a copy of the attached advisory and that you understand the information regarding the proper usage of the Decompressor. Once the form has been signed by all physicians, the form must be completed and signed by the recall coordinator. Please fax to 866-521-2762 or email to [email protected].
- Quantity in commerce
- 20,385 total distributed since 2004
- Distribution
- Nationwide Distribution including the states of AR, AZ, CA, CO, FL, GA, IL, KS, MS, ND, and NE.
- Product codes and lots
- Part Number(s) 0407-251-000, 0407-260-000, 0407-265-000, 0407-266-000, 0407-280-000, and 0407-281-000.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0127-2014 | Class II | Terminated | Voluntary: Firm initiated |