Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids, or in...
Ethicon Endo-Surgery Inc began a recall of Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids, or in various kit configurations with one band per kit... on . FDA classified it as Class II and gives the reason as "During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0149-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids, or in various kit configurations with one band per kit. Product Usage: Intended for use in surgical treatment of morbid obesity
- Product code
- LTI Implant, Intragastric For Morbid Obesity
- Reason for recall
- During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- Ethicon sent an Urgent: Medical Device Recall letter dated September 30, 2013 to affected consignees via overnight, UPS to allow tracking of the receipt. The letter identified affected products, problem and actions to be taken. Each consignee is requested to respond to a provided Business Reply Form (BRF), indicating whether they have affected product, how much and whether they will/have returned the product or whether they no longer have any of the recall product. For questions contact your local sales representative or call our Customer Support Center 1-877-384-4266 option 6.
- Quantity in commerce
- Product codes: RLZB22 - 2,352 units; RLZB22D1 - 968 units; RLZB22DG1 - 327 units; RLZB22DGT - 20 units; RLZB22DT - 48 units; RLZB22T - 0 units
- Distribution
- Worldwide Distribution: USA (nationwide) including the states of: Alabama, Connecticut, Indiana, Michigan, North Carolina, South Carolina, Alaska, Delaware, Kansas, Minnesota, Ohio, Tennessee, Arizona, Florida, Kentucky, Mississipp,i Oklahoma, Texas, Arkansas, Georgia, Louisiana, Nebraska, Oregon, Virginia, California, Idaho, Maryland, Nevada, Pennsylvania, Washington, Colorado, Illinois, Massachusetts, New Jersey, Rhode Island, and West Virginia; and the countries of: Argentina, France, Malaysia, Spain, Australia, Georgia, Martinique, Sweden, Belgium, Germany, Mexico, Switzerland, Brazil, Greece, Netherland, Taiwan, Chile, Hungary, Panama, Tunisia, China, India, Philippines, United Arab Emirates, Czech Republic, Israel, Poland, United Kingdom, Egypt, Italy, Portugal, Venezuela, Estonia, Kenya, Russia, Finland, Kuwait, and Saudi Arabia.
- Product codes and lots
- The following product code and lot numbers were affected by this recall: Product Codes: RLZB22 Lot Numbers: ZJKBK6, ZJLBBR, ZJLBBT, ZJLBBV, ZJMBB5, ZJMBCG, ZJMBCH, ZJMBCJ, ZJMBCK, ZJMBCL, ZJMBJC, ZJMBJD, ZJMBJF, ZJPBD1, ZJPBD2, ZJPBFT, ZJPBFV, ZKBBCC, ZKBBCD, ZKBBCF, ZKBBCG, ZKBBGC, ZKBBGD, ZKBBGJ, ZKBBJN, ZKBBNV, ZKBBNW, ZKBBNY, ZKCBBK, ZKCBDD, ZKCBJJ, ZKMBK4, ZKMBKP, ZKNBBB, ZKNBCL, ZLBBCH, ZLBBFJ, ZLBBFL, ZLCBCB, ZLCBD8, ZLDBCZ, ZLDBDM, ZLDBDV, ZLFBF7, ZLFBFB, ZLLBCR, ZLLBFT, ZLMBCH, ZLMBWD, ZLNBBF, ZLNBGZ, ZMCBDH, ZMDBCH, ZMFBC4, ZMGBBF, ZMGBC1, ZMHBBW, ZMJBHV, ZMKBBH, ZMLBCJ, ZMNBCK, ZNBBD2, ZNFBCC, ZNHBCM, and ZNLBB4. Product Codes: RLZB22D1 Lot Numbers: G4TA93, G4TC3R, G4TD5R, G4TF1X, G4TG22, G4TK21, G4TL1D, G4TM3L, G4TM84, G4TP0W, G4TT4D, G4TU2V, G4TV3T, G4TW6D, G4U17U, G4U75W, G4U795, G4UD6T, G4UF6Y, G4UG4E, G4UH49, G4UH4K, H43A3N, H43C8P, H43E9V, H43J2P, H43N12, H43N42, H43P5U, H43P9C, H43T3F, H43U78, H43V8J, H43X4K, H43Z8Y, H4453N, H44623, H4478W, H44965, H44C9G, H44D8W, H44K3G, H44K9M, H44N0X, H44R1Y, H44R4U, H44T9D, H44V39, H44W6H, H44Y24, H44Z0Y, J4A04D, J4A11V, J4A274, J4A668, J4A81X, J4A96M, J4AC6L, J4AE3P, J4AH2Z, J4AK89, J4AN34, J4AU9R, J4AW1Y, J4C35K, J4C76E, J4CE9V, K4C518, K4CE4L, K4CT8W, K4D17T, and K4D45Y. Product Codes: RLZB22DG1 Lot Numbers: G4TH90, G4TJ61, G4TP6J, G4TR76, G4TV48, G4TX1V, G4U184, G4U734, G4UD77, G4UF9V, G4UH6P, H43A3P, H43C8R, H43G7Y, H43T3K, H43U7E, H43X1G, H43Z5Y, H44645, H44H9M, H44K2T, H44L44, H44V3A, H44W6G, H44Y25, H44Z0X, H44Z3N, J4A48K, J4AA11, J4AL00, J4AR9K, and K4CE6X . Product Codes: RLZB22DGT Lot Numbers: G4U323, J4AY55, and K4C663. Product Codes: RLZB22DT Lot Numbers: G4UD92, H44D1Y, J4AY72, and K4C666. Product Codes: RLZB22T Lot Numbers: G4RE82
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0149-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0150-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0151-2014 | Class II | Terminated | Voluntary: Firm initiated |