510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0202-2014Class II

Synthes 8.5 MM Medullary Reamer

Synthes GmbH / initiated 2013-10-03 / Terminated / root cause: design

Synthes USA HQ, Inc. began a recall of Synthes 8.5 MM Medullary Reamer Product Usage: An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to... on . FDA classified it as Class II and gives the reason as "Complaints were received for intraoperative reamer head breakages which can allow for unretrieved fragments of non-implant grade material". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0202-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes 8.5 MM Medullary Reamer Product Usage: An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, bmT, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Complaints were received for intraoperative reamer head breakages which can allow for unretrieved fragments of non-implant grade material.
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent an Urgent Notice: Voluntary Medical Device Recall Notification letters, dated October 3, 2013, to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their inventory and immediately remove the affected lots from stock. For questions call 610-719-5450 or contact your Synthes Sales Consultant.
Quantity in commerce
2979
Distribution
Worldwide Distribution: USA (Nationwide) and countries of: Canada and Mexico.
Product codes and lots
Part No. 352.085 with lot numbers 20141 through 24510, and 4399543 through 5870257

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0202-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0202-2014Class IITerminatedVoluntary: Firm initiated