Synthes 8.5 MM Medullary Reamer
Synthes USA HQ, Inc. began a recall of Synthes 8.5 MM Medullary Reamer Product Usage: An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to... on . FDA classified it as Class II and gives the reason as "Complaints were received for intraoperative reamer head breakages which can allow for unretrieved fragments of non-implant grade material". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0202-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes 8.5 MM Medullary Reamer Product Usage: An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, bmT, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Complaints were received for intraoperative reamer head breakages which can allow for unretrieved fragments of non-implant grade material.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Synthes sent an Urgent Notice: Voluntary Medical Device Recall Notification letters, dated October 3, 2013, to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their inventory and immediately remove the affected lots from stock. For questions call 610-719-5450 or contact your Synthes Sales Consultant.
- Quantity in commerce
- 2979
- Distribution
- Worldwide Distribution: USA (Nationwide) and countries of: Canada and Mexico.
- Product codes and lots
- Part No. 352.085 with lot numbers 20141 through 24510, and 4399543 through 5870257
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0202-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY