Z-0968-2014Class II

VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems

Ortho-Clinical Diagnostics, Inc. / initiated 2013-10-10 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate... on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics issued a product correction notification for VITROS Chemistry Mixing Cup Arrays utilized on VITROS 250 and VITROS 350 Chemistry Systems. Increase in Software error due to affected Mixing Cup...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0968-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate container used in making automatic (auto)-dilutions on the VITROS 250/350 Chemistry Systems. The VITROS 250 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Microslides. The VITROS 350 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides.
Product code
JRB Micro Mixer
Reason for recall
Ortho Clinical Diagnostics issued a product correction notification for VITROS Chemistry Mixing Cup Arrays utilized on VITROS 250 and VITROS 350 Chemistry Systems. Increase in Software error due to affected Mixing Cup Arrays.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification dated October 10, 2013 to all affected customer. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to complete and return the attached Confirmation of Receipt form. For questions contact the Customer Technical Services at 1-800-421-3311.
Quantity in commerce
6380 units (each unit contains 48 arrays with 10 microcups)
Distribution
Worldwide Distribution - US Nationwide and the countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
Product codes and lots
1) VITROS 250 (Product Code 8132086; Model # 3332); 2) VITROS 350 (Product Code 6802153; Model # 3330); 3) Mixing Cup Arrays (Product Code 1631779; Model # n/a)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0968-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0968-2014Class IITerminatedVoluntary: Firm initiated