Z-0519-2014Class II

Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device...

Siemens Healthcare Diagnostics Inc. / initiated 2013-10-02 / Terminated

Beckman Coulter Inc. began a recall of Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0519-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I
FDA root cause
Pending
Action
Beckman Coulter sent an Urgent Product Correction (PCA) letter dated October 1, 2013, along with a customer response form to all affected customers on October 2, 2013. The customer letter informed customers of the recall with the issue, impact, action (If your Access Immunoassay System experiences Wash Carousel Motion Errors that cannot be corrected with normal troubleshooting measures as outlined in the Access 2 Reference Manual section 6.3, contact Beckman Coulter Customer Technical Support. If you do not have Access System RVs with the lot numbers listed above, or if your Access Immunoassay System is not experiencing Wash Carousel Motion Errors, no additional action is required.), resolution, and contact information, website: http://www.beckmancoulter.com. Customers in the United States and Canada with questions were instructed to call 1-800-854-3633, outside the United States and Canada, customers were instructed to contact their local Beckman Coulter representataive. For questions regarding this recall call 714-961-4941.
Quantity in commerce
16,154 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, India, Hong Kong, Guatemala, China, Brazil, Brunei Darussalam, Costa Rica, El Salvador, JordanMalaysia, Mexico, Panama, Philippines, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Venezuela, and Viet Nam.
Product codes and lots
Part Number 81901, Lot Numbers: 13181168-13341170.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0519-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0519-2014Class IITerminatedVoluntary: Firm initiated