Z-0429-2014Class II
Synthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the...
Synthes GmbH / initiated 2013-08-01 / Terminated
Synthes USA HQ, Inc. began a recall of Synthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereosteal cavity for percutaneous plate insertion on . FDA classified it as Class II and gives the reason as "The small extendible Soft Tissue Retractor was recalled due to locking nut (clamping sleeve) malfunction". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0429-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereosteal cavity for percutaneous plate insertion.
- Product code
- GAD Retractor
- Reason for recall
- The small extendible Soft Tissue Retractor was recalled due to locking nut (clamping sleeve) malfunction.
- FDA root cause
- Use error
- Action
- Synthes sent an Field Safety Notification letter dated August 1, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to take the following actions: Exchange the old surgical technique guide with the new guide version F provided with this notification. Review the revised surgical technique guide. Forward this Field Safety Notification to anyone in your facility that needs to be informed. If the technique guide has been forwarded to another facility, contact that facility. Maintain awareness of this notice until all technique guides have been exchanged. Maintain a copy of this notice. Please complete and return this page by either of the following methods: Fax: 877-567-9320 or Scan/email: [email protected] Thank you for your attention to this issue.
- Quantity in commerce
- 374
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada and Mexico.
- Product codes and lots
- Part J5957AC (old), and Part J5957F (new)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0429-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.