Aesculap GNI161 Bipolar Foot Control
Aesculap, Inc. began a recall of Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar... on . FDA classified it as Class II and gives the reason as "Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0479-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Aesculap sent an Important Recall Notification letter dated September 23, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to return affected product and receive replacement product or credit. An Inventory Sheet is attached and must be completed. For questions call 610-984-9265 or 610-984-9414.
- Quantity in commerce
- 85
- Distribution
- Worldwide Distribution - USA Nationwide and countries of: Austria, Germany, Japan, and Taiwan.
- Product codes and lots
- all GN161's distributed between 4/28/2009 and 5/22/2013.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0479-2014 | Class II | Terminated | Voluntary: Firm initiated |