Resolute Integrity Zotarolimus-eluting Coronary Stent: Stent ID 3.0 mm; Stent length 34 mm; Rx, Sterile. Mfg...
Medtronic Vascular, Inc. began a recall of Resolute Integrity Zotarolimus-eluting Coronary Stent: Stent ID 3.0 mm; Stent length 34 mm; Rx, Sterile. Mfg in Medtronic Ireland, Galaway, Ireland... on . FDA classified it as Class II and gives the reason as "A small number of units in lot number 0006573585 of the Resolute Integrity Zotarolimus-Eluting Coronary Sten System may contain a compliance chart incorrectly referencing a 2.25 mm diameter instead of the correct 3.0mm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0154-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Resolute Integrity Zotarolimus-eluting Coronary Stent: Stent ID 3.0 mm; Stent length 34 mm; Rx, Sterile. Mfg in Medtronic Ireland, Galaway, Ireland. Distributed by: Medtronic Inc. Minneapolis, MN. Indicated for improving coronary luminal diameters in patients.
- Product code
- NIQ Coronary Drug-Eluting Stent
- Reason for recall
- A small number of units in lot number 0006573585 of the Resolute Integrity Zotarolimus-Eluting Coronary Sten System may contain a compliance chart incorrectly referencing a 2.25 mm diameter instead of the correct 3.0mm.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Urgent Medical Device Recall Letters were sent on October 7, 2013 via UPS 3-day service.
- Quantity in commerce
- 201 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Model: RSINT30034UX; Customer Facing Number: RSIST30034UX. Expiration date: 25 May 2014.