Z-0455-2014Class II

Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel...

Synthes GmbH / initiated 2013-09-30 / Terminated / root cause: manufacturing process

Synthes USA HQ, Inc. began a recall of Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody... on . FDA classified it as Class II and gives the reason as "A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0455-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1.
Product code
OVE Intervertebral Fusion Device With Integrated Fixation, Cervical
Reason for recall
A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Synthes sent an email message dated March 23, 2012, to the sales representative who received the affected product. The email identified the product, the problem, and the action to be taken. The letter requested the sales rep examine inventory, stop using the product, and quarantine the affected product to ensure that it will not be used. Synthes instructed the customer not to remove any product from their account or return any product without further instruction. For questions they were instructed to call 610-719-5364. For questions regarding this recall call 610-719-5000.
Quantity in commerce
1
Distribution
Nationwide Distribution - only AZ .
Product codes and lots
part no. 04.647.117S, Plate implant no. 7706771, spacer lot no. 7675745

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0455-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0455-2014Class IITerminatedVoluntary: Firm initiated