Z-0099-2015Class II
CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"
Intra Pump Infusion Systems / initiated 2013-10-18 / Terminated / root cause: design
Intra Pump Infusion Systems began a recall of CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle" on . FDA classified it as Class II and gives the reason as "Product has the potential to have a clear debris on the tip of the needle". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0099-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intra Pump Infusion Systems
- Product
- CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"
- Product code
- FRN Pump, Infusion
- Reason for recall
- Product has the potential to have a clear debris on the tip of the needle.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Cane S.p.A. Medical Technology sent an Urgent: Correction and Removal Action Required letter dated October 21, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: a. Stop using the 19G x 11/2" 1.10 x 40mm steelneedle included in the CRONO Syringe blister pack. B. Safely discard the 19G x 11/2" 1.10 x 40mm steel needle into a Sharps container. c. Use a TRANSFER DEVICE (e.g.,Needleless adapter with female-to-female luer connector) or alternative device to transfer medications into the CRONO syringe. d.For intravenous infusions,we are recommending that a filter also be used during infusions. e. Complete the required 'Acknowledgment and Receipt Form'and fax back to our Distributor;Intra Pump Infusion Systems at 630-845-2768. Contact Information for our US Distributor: INTRA PUMP INFUSION SYSTEMS " Monday through Friday 8:00am to 5:00pm CST,TollFree:1-866- 211-7867 " Direct:630-845-7867 " Fax:630-845-2768 www.lntrapump.com
- Quantity in commerce
- 25,440 pieces
- Distribution
- Worldwide Distribution: US (nationwide) in states of CA, PA, MA, IL, OH and NY; and Internationally to Canada.
- Product codes and lots
- Product Code CRN20, lot numbers 14N02, 15E01, 15A02
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0099-2015 | Class II | Terminated | Voluntary: Firm initiated |