Z-0099-2015Class II

CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"

Intra Pump Infusion Systems / initiated 2013-10-18 / Terminated / root cause: design

Intra Pump Infusion Systems began a recall of CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle" on . FDA classified it as Class II and gives the reason as "Product has the potential to have a clear debris on the tip of the needle". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0099-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intra Pump Infusion Systems
Product
CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"
Product code
FRN Pump, Infusion
Reason for recall
Product has the potential to have a clear debris on the tip of the needle.
Root cause tag
design
FDA root cause
Device Design
Action
Cane S.p.A. Medical Technology sent an Urgent: Correction and Removal Action Required letter dated October 21, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: a. Stop using the 19G x 11/2" 1.10 x 40mm steelneedle included in the CRONO Syringe blister pack. B. Safely discard the 19G x 11/2" 1.10 x 40mm steel needle into a Sharps container. c. Use a TRANSFER DEVICE (e.g.,Needleless adapter with female-to-female luer connector) or alternative device to transfer medications into the CRONO syringe. d.For intravenous infusions,we are recommending that a filter also be used during infusions. e. Complete the required 'Acknowledgment and Receipt Form'and fax back to our Distributor;Intra Pump Infusion Systems at 630-845-2768. Contact Information for our US Distributor: INTRA PUMP INFUSION SYSTEMS " Monday through Friday 8:00am to 5:00pm CST,TollFree:1-866- 211-7867 " Direct:630-845-7867 " Fax:630-845-2768 www.lntrapump.com
Quantity in commerce
25,440 pieces
Distribution
Worldwide Distribution: US (nationwide) in states of CA, PA, MA, IL, OH and NY; and Internationally to Canada.
Product codes and lots
Product Code CRN20, lot numbers 14N02, 15E01, 15A02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0099-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0099-2015Class IITerminatedVoluntary: Firm initiated